TOPTICAs proportional integral differential regulator PID 110

PID 110 – Versatile PID Regulator

  • Ideally suited for side-of-fringe laser stabilization
  • P, I, D parameters individually adjustable
  • Variable total gain
  • High Voltage output e.g. for driving piezos
  • Automatic ReLock feature

The PID 110 is a proportional integral differential regulator for the SYS DC electronics system and can be used for various control tasks, such as stabilization of the laser frequency to the edge of an atomic resonance or a transmission of a cavity. Both high and low voltage actuators (Piezos, laser temperature, …) are supported. Its flexibility makes the PID 110 into a general purpose PID for scientists laboratories.

Technical Information

Technical Information

 

Block diagram of PID 110

 

Specifications PID 110:

  • Side-of-fringe stabilization
  • Regulator output -5  .. +150 V, max 30 mA
  • External modulation input -5 V .. +5 V
  • LV small signal bandwidth up to 1.5 MHz
  • Gain * bandwidth product 3 MHz (for gains ≥ 2)
  • HV small signal bandwidth > 10 kHz (current limited to 30 mA)

 

Accessible Parameters:

  • P, I, D -contribution of the control loop can be adjusted individually by precision trimmer
  • Presetable set point for the control loop, e.g. stabilization to side-of-fringe
  • Useful optical display of the control lock status with 3 LEDs (search, unlock, reset)
  • Ramp/Modulation ON/OFF switch and regulation ON/OFF switch
  • Inversion of the control output with switch POS/NEG
  • On request a relock-mechanism can be activated (automatic search for the next stabilization mode) when stabilizing on external resonators, FPI´s etc., indicated by red relock-LED
  • External modulation input or internal synchronization with scan control SC 110 possible
  • Controller ON/OFF switch
  • HF-tight cassette, width 12 TE (60 mm)

 

Input / Outputs:

  • SubD9 connector for actual signal in, controller output voltage
  • BNC connectors for external modulation input
  • BNC connector for buffered monitor signal, modulation, HV/100, controller output, error signal, signal in, set point
  • BNC connector for regulator output: -5...150 V, max 30 mA

Options & Related Products

Options & Related Products

 
Related Products:

PID 110LIR 110PDD 110FALC 110DigiLock 110LaseLockWavelength MeterIscan
Description / typeAnalog
PID
Analog
Lock-In
Regulator
Pound-Drever-
Hall Signal
Generation
Fast Analog
2-Channel
PID
Versatile Digital
Locking Solution
Suited for
third party
lasers
Wavelength meter with PID optionLow finesse
etalons as reference with
Lock-Electronics
Side-of-fringe
Not applicable

Top-of-fringeWith PDDWith PDD
Locking bandwidth*kHz .. MHzkHz range≈ 45 MHz≈ 10 MHz1 MHz< 100 HzkHz range
Modulation frequency0.6 Hz .. 14 kHz20 MHz **17 Hz .. 25 MHz33 Hz .. 1 MHzNot applicable
AccuracyDepends on reference (can be very good)> 10 MHz1 MHz (relative)
Signal analysis
Relock mechanisms
Computer control
High voltage output
High bandwidth output
Two channel version
SYS DC 110 module
Stand-alone
low budgethigh end & preferrable

*Estimated bandwidth depends on gain and PID settings  **5- 40 MHz versions available

Technical Drawings

Technical Drawings

Applications

Applications

Laser locking and linewidth narrowing
toptica_AP_1012_laser_locking_2012.pdf
Trapping and quantum computing
toptica_AP_1004_quantum_computing_2007.pdf

Download

Download

Brochure: Laboratory Electronics
toptica_BR_SC_Laboratory_Electronics.pdf
Catalog: Product Catalog Research Grade Lasers (10 MB)
toptica_BR_product_catalog.pdf